1
|
- Adriano Farina
- Liceo Ginnasio statale G.D.
Romagnosi, Parma, Italy
- adriano@pcfarina.eng.unipr.it
|
2
|
- Measuring noise exposition of teenager subjects with real-life volume
settings and complying with the following:
- Following international and european standards about measurements
techniques
- Using a test signal that is both standard and similar to real-life music
|
3
|
|
4
|
- One minute of pink noise was generated
- It was equalized in order to obtain the desired spectrum
- The resulting sound had an average RMS value of -16dBFS instead of the
standard -10
|
5
|
- In order to obtain the correct RMS value we employed the Hard Limiting
module of Adobe Audition, with a 6dB gain
- The Graphic Equalizer was again used to correct the minor distortion
caused by the Hard Limiting
- The resulting sound was measured compliant with the IEC standard
|
6
|
- However, it was noted by Alastair Hardie, a Senior Electronic Engineer
for Frontier Silicon, that the Crest Factor had a 3.1373 / 3.1372 ratio,
instead of the 1.8/2.2 specified in section 5.1 of standard EN
50332-1:200
|
7
|
- While the IEC standard states that the programme simulation noise must
have a crest factor ranging between 1.8 and 2.2, it was technically
impossible to generate such a noise. In fact the standard is written
considering an analog crest factor measurement.
|
8
|
- Analog measurements were performed using a quasi-peak detector, which
systematically underestimated the result.
- It was thence attempted to emulate it employing the Statistical Analysis
tool of Adobe Audition.
- Specifying a window width of 35 ms, this tool computes correctly the
pseudo-peak value as maximum RMS, if you add 3dB to the result (or
by specifying that 0 dB = FS sine wave). It was checked that with these
settings one gets the readings specified in table A-II of the IEC
standard, employing a 5kHz tone burst of 1ms length.
|
9
|
|
10
|
|
11
|
|
12
|
|
13
|
- First we had to measure an average spectrum from the collected music,
which was done with the following system:
|
14
|
|
15
|
- We then proceeded to generate the signal using the same method employed
for the IEC one
|
16
|
|
17
|
|
18
|
- In order to perform the measurements the test signals were put on the
digital audio players using the best codec available for each device
(uncompressed wav when available). The formats employed are the
following:
- Uncompressed WAV (44100 Hz, 16 bits, stereo)
- WMA Lossless
- WMA 192 kbps
- WMA 128 kbps
- MP3Pro 144 kbps
- MP3 192 kbps
- MP3 128 kbps
- Apple Lossless
- AAC 192 kbps
- AAC 128 kbps
|
19
|
|
20
|
- This dummy head is compliant with IEC 60959, and is specifically
manufactured for testing “internal” hearing aids
- The microphones have been calibrated, removing the pinna, and inserting
over the capsule a Bruel & Kjaer type 4100 reference sound source,
which provides a pure tone at the frequency of 1 kHz and with an RMS
sound pressure level of 94 dB (1 Pa).
- The calibration signal was employed for setting the calibration of the
SpectraRTA program, as shown here:
|
21
|
- However, it is also necessary to correct for the frequency response of
this specific Head and Torso Simulator. The manufacturer does not
provide a suitable free-field frequency response for the Ambassador
dummy head. So it was necessary to employ the results of anechoic
impulse response measurements which had been previously performed on
this specific dummy head at the anechoic chamber of Winterthur
(Switzerland), kindly made available by Rieter Automotive.
|
22
|
- SpectraRTA already provides the capability of correcting for the
frequency response of the microphones employed, so the compensation of
the frequency response of the Ambassador dummy head did not require any
effort.
|
23
|
- SpectraRTA was configured for measuring a linearly-averaged spectrum in
1/3 octave bands, averaging the signal of both channels (ears), and
computing an unweighted spectrum and an A-weighted wideband value. Each
test signal was 60 s long, but the measurement time was set 30 s,
leaving 20s at beginning for allowing the device to stabilize before
starting the measurement.
|
24
|
- Each device was measured 5 times, dismounting and remounting the
headphones each time, as recommended by standard EN 50332, in order to
reduce the mounting error. The results were then averaged.
- For ensuring a better correlation between the results and in order to
minimize the measurement time, we used the same headphone position with
both test signals (IEC and MUSIC). The headphone was inserted, then the
two signals were measured and then the headphones were removed and
reinserted.
- The volume control of the player was left untouched since the last usage
from the owner of the device. These results are thence not significant
for discriminating "dangerous" devices from "safe"
devices.
|
25
|
|
26
|
|
27
|
|
28
|
- For each device under test it was possible to obtain two values of the
“exposure sound pressure level”: the first based on the IEC programme
test signal, the second on the MUSIC test signal. The following table
shows the results, in terms of average SPL +/ the standard deviation.
|
29
|
|
30
|
- Results are usually quite similar for the same device
- Huge differences between devices
- No difference between the signals
|
31
|
|
32
|
- At 3150 Hz a strong ear duct resonance is present, due to the air
trapped behind the ear bud.
- Looking at the free-field frequency response of the Ambassador dummy
head, the peak in the frequency response was instead at 5 kHz,
corresponding to the “dip” in the curves of the figure.
|
33
|
|
34
|
|
35
|
- It is wrong to employ the free-field frequency response for correcting
the recorded signals, as the free-field response does not take into
account the modification of the ear duct resonance occurring when an ear
bud is inserted in the pinna.
- This can be seen as a severe inconsistency of the current EN 50332-1
standard.
- It could be more advisable to employ a diffuse-field response (which is
usually smoother)
|
36
|
- A significant number of the devices under test was found set for over 90
dB(A) playback levels.
- In terms of assessment of noise-induced health risk, usually a “safe”
value is considered to be a daily exposure of 8 h at 80 dB(A).
- Whenever higher SPL values are present, the duration of the exposure
should be reduced, in order of keeping the same daily “noise dose”. An
energetic equivalence principle is assumed, which means that the
exposure should be reduced at 4h for an SPL of 83 dB(A), to 2h for an
SPL of 86 dB(A), and so on.
- For each of the devices under test, it was computed what is the maximum
time allowed daily for employing it for listening to music, as shown in
the following table:
|
37
|
|
38
|
- 5 of the 13 devices have a time limit below 1h
- Most users declared a 1 to 2 hours average usage
- A proper hearing risk assessment should consider a whole day, with all
its activities.
- A lot of money has been spent for reducing the noise at the workplace,
but this will be wasted if we allow our teenagers to be over-exposed for
years due to the usage of personal digital audio players capable of
dangerous levels.
|